KPR#:1152274  Product:    1.0

Keyword: 

One-line Description:
Task Scheduling Discrepancy While Choosing Days of the Week

Problem:

When scheduling tasks, the instrument chooses the day of the week based on the UTC time zone. If the user schedules a time of day in their time zone that has a different day of the week in the UTC time zone, the task will run at the time scheduled, but on the wrong day of the week.

Temporary Solution:

When scheduling tasks, compare the time zone of the instrument with the UTC time zone to determine which day of the week the task will run.

Fix Information:

n/a


KPR#:1134212  Product:    1.0

Keyword: 

One-line Description:

Page refresh required when turning off HTTPS/TLS

Problem:

When HTTPS/TLS is active in the Assist Control Software and then turned off via Settings - Security - HTTPS/TLS, it is no longer possible to log into the Assist Control in the same browser session from external devices.

The Assist Control Interface not affected.

Temporary Solution:

Restart the browser

Fix Information:

n/a


KPR#:1123396  Product:    1.0

Keyword: 

One-line Description:

Run Control shows different run times on Assist Interface and in PC browser

Problem:

When comparing the task (e.g. Make Ready) run times in the Run Control between the Assist Interface at the instrument and in a PC browser window, the run times are not identical.
This happens when the times on the Assist Hub or the time on the PC are not correctly set - most likely because Assist Hub Date & Time Source is set to "Manual"

Temporary Solution:

Ensure:
- PC correctly receives time from a Time Server
Check
- in Assist Control Software - Settings - General - Date & Time: Set Date & Time Source to NTP (Network Time Protocol) when the Assist Hub can reach a time server in the network/internet
- in Assist Control Software - Settings - General - Date & Time: Set Date & Time Source to CDS when LC Driver with revision >= 3.8 is installed in the CDS - the driver will then provide the time from the CDS server PC

Fix Information: