KPR#:960279  Product:    1.0

Keyword: 

One-line Description:
User with restricted permission in CDS can modify in Assist Control Softare

Problem:

A user which can log into a CDS with his user account and log into the Assist Control Software with one of the available roles may be able to modify method parameters (e.g. Flow) in the Assist Control Software, despite not being permitted to in the CDS.

Temporary Solution:

CDS permissions and the Assist Control Software are independent.
If the users role on the Assist Control Software has to many permissions, assign a proper role with less permissions.

Fix Information:

n/a


KPR#:1101923  Product:    1.0

Keyword: 

One-line Description:
Search in context help may fail on Assist Control Interface

Problem:

The search in the context help might fail when executed on the Assist Control Interface

Temporary Solution:

n/a

Fix Information:

Fixed in 2.0


KPR#:1134212  Product:    1.0

Keyword: 

One-line Description:
Page refresh required when turning off HTTPS/TLS

Problem:

When HTTPS/TLS is active in the Assist Control Software and then turned off via Settings - Security - HTTPS/TLS, it is no longer possible to log into the Assist Control in the same browser session from external devices.

The Assist Control Interface not affected.

Temporary Solution:

n/a

Fix Information:

Fixed in 2.0


KPR#:1143678  Product:    1.0

Keyword: 

One-line Description:
Assisted Troubleshooting may fail after rebooting detector module

Problem:

Assisted Troubleshooting of a "Lamp Ignition Failed" error may fail during Verification step when the detector module had been turned off as part of the "Replacing the lamp" procedure. Abort reason is "Attempt to launch a sub procedure..."

Temporary Solution:

Workaround: When Verification step failed, cancel the Assisted Troubleshooting, navigate to Status screen, use the "Switch On Lamp" quick action and navigate to Health - Diagnostics - Intensity test to manually verify the new lamp.

Fix Information:

Fixed in 2.0


KPR#:1239180  Product:    1.1

Keyword: 

One-line Description:
Make Ready / Standby flush steps limiting run time

Problem:

During execution of Make Ready or Standby Task, the steps Pre-Flush, Flush System and Store Column will complete after one hour when configured flush volume and pump flow result in a flush time greater than 1h.
The time limit is not visible in the configuration UI or the generated result.

Temporary Solution:

Ensure the configured run time resulting of flush volume and pump flow do not exceed 1h.

Fix Information:

Fixed in 2.0 by limiting the maximum flush time to 3 h. 