Resolved in Version: 2.2

Preface
-------------------------
This Software Release Bulletin documents all fixes and enhancements that are incorporated in the new release.
-------------------------


Keyword Glossary




Keyword Index



KPR#:1277576  Product:    2.1

Keyword: 

One-line Description:
Scheduled Make Ready/Standby may not start when multiple schedules defined

Problem:

Make Ready or Standby tasks may not be started at the scheduled time, when multiple schedules have been defined and partly deleted.

Temporary Solution:

Delete schedules for affected tasks and duplicate the task. Then create a new schedule for the new task.

Fix Information:

Fixed in version 2.2


KPR#:1308309  Product:    2.1

Keyword: 

One-line Description:
Make Ready Task not switching G1170A positions when 2D-LC Cluster is active

Problem:

When running "Make Ready" tasks on an instrument configured as 2D-LC Cluster, the configuration UI allows to select a valve position for the 2D valves. However, during execution (when active pump is the 1D pump) those valves are not switched due to the active cluster.

Temporary Solution:

When configuring the "Make Ready" task ignore the 2D valves in the Flow Path section of the Equilibration step.

Fix Information:

Fixed in 2.2


KPR#:1308435  Product:    2.1

Keyword: 

One-line Description:
Make Ready/Standby tasks invalid when configured pump / thermostat / detector exchanged

Problem:

When exchanging HW module (pump / thermostat / detector) which is configured in Make Ready / Standby tasks, tasks get invalid and user must manually exchange the configured pump etc. in each task

Temporary Solution:

Manually reconfigure each invalid task depending on the new instrument setup

Fix Information:

Fixed in 2.2


KPR#:1364699  Product:    2.1

Keyword: 

One-line Description:
Assist Hub clock reset not shown in log

Problem:

Assist Hub may loose it's time after being powered down for longer time. After boot the Assist will get a new time set depending on configuration (NTP, CDS Synchronization, Manual). However, the Log does not show an appropriate log entry for the time reset

Temporary Solution:

n/a

Fix Information:

Fixed in 2.2